LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
CO2 Adapter Cable (W30) Replacement
Section Menu Section Contents Back Index
354
8
Removing the CO2 Adapter Cable (W30)
Refer to Figure 8.119 on p. 353.
To remove the CO2 adapter cable (3012397-01) from the rear case:
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in Removing the System (A01)/Therapy (A04) PCB Assembly (p. 233).
3. Remove the OEM PCB (A06) as described in Removing the OEM PCB (A06) (p. 262).
4. Remove the NIBP/CO2 module as described in Removing the NIBP (A21)/CO2 (A23) Modules (p. 288).
5. Disconnect the CO2 adapter cable from the CO2 PCB (A23) at J1.
Installing the CO2 Adapter Cable (W30)
Refer to Figure 8.119 on p. 353.
To install the CO2 adapter cable (3012397-01) from the rear case:
1. Connect the CO2 adapter cable to the A23 CO2 PCB at J1.
2. Install the NIBP/CO2 Module as described in Installing the NIBP (A21)/CO2 (A23) Modules (p. 292).
3. Install the OEM PCB (A06) as described in Installing the OEM PCB (A06) (p. 265).
4. Install the system/therapy PCB assembly as described in Installing the System (A01)/Therapy (A04) PCB Assembly (p. 239).
5. Reassemble the case as described in Reassembling the Case (p. 184).