7 USER’S STATEMENT Chapter A: GENERAL & TECHNICAL DATA
16 Planmeca ProOne Technical manual
7 USER’S STATEMENT
Radiation leakage technique factors
The maximum rated peak tube potential is 70 kV and the maximum rated continuous tube
current is 2 mA for the maximum rated peak tube voltage.
Filtration
The radiation port contains additional filtration of at least 1.5 mm aluminium.
Total filtration min. 2.5 mm Al.
When the X-ray beam is attenuated with 1.8 mm Al the resulting dose is 0.5 - 0.7 times the
original.
Maximum attenuation equivalent of the front panel of the panoramic
sensor head
0.5 mmAl
Rated line voltage
100 - 240 V~
Maximum line current
Maximum 8 Amperes at 100 V~, 4A at 230 V~
Technique factors that constitute the maximum line current
condition
70 kV / 7 mA at 100 V~
Generator rating and duty cycle
SW controlled at least 1:20 at 25°C ambient temperature with 70 kV, 7 mA, 10 s.
Maximum deviation of peak tube potential from indicated value
±5%
Maximum deviation of tube current from indicated value
±10%
Maximum deviation of exposure time from indicated value
±10%

Table of Contents

Other manuals for Planmeca ProOne

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProOne and is the answer not in the manual?

Planmeca ProOne Specifications

General IconGeneral
BrandPlanmeca
ModelProOne
CategoryMedical Equipment
LanguageEnglish

Related product manuals