Chapter C: ERROR MESSAGES 1 ERROR MESSAGES
Technical manual Planmeca ProOne 83
1.5 System related (E06-XXX)
1.6 Calibration related (E07-XXX)
Code Explanation Comments
E06-005 PSU High temperature PSU temperature is too high.
Wait for cooling.
E06-011 Tube Low temperature Tube temperature is too low.
Wait for warming. If necessary, replace the
tube head/ PSU PCB. Then perform the
filament definition again.
E06-012 Open circuit Tube temperature measurement circuit is
open.
Replace the tube head and perform the
filament definition again.
E06-013 PSU Long exposure time Exposure time is too long.
Replace the PSU assembly and perform the
filament definition again. If necessary, then
replace the CPU PCB.
E06-016 Exposure switch Exposure button Exposure button is stuck or continuously
pressed down.
If necessary, replace the exposure switch.
E06-017 PSU Temperature out of limits PSU temperature is out of bounds.
Replace the tube head.
Code Explanation Comments
E07-001 Tube Filament definition X-ray tube filament definition has not been
performed.
Perform the filament definition.
E07-005 Dimax sensor Calibration The Dimax sensor has not been calibrated.
Calibrate the sensor.
E07-006 Tube Filament definition X-ray tube filament definition was not
successful.
Repeat the filament definition. If necessary,
then replace the tube head.

Table of Contents

Other manuals for Planmeca ProOne

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Planmeca ProOne and is the answer not in the manual?

Planmeca ProOne Specifications

General IconGeneral
BrandPlanmeca
ModelProOne
CategoryMedical Equipment
LanguageEnglish

Related product manuals