EasyManua.ls Logo

Planmeca ProOne - Patient Positioning for Multi-angle TMJ Exposures

Planmeca ProOne
195 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter B: SETTINGS 4 TECHNICAL SETTINGS
Technical manual Planmeca ProOne 57
4.1 Technical: Mode (T3100)
To select the operating mode of the X-ray unit, first select the option Mode (T3100) on the
Technical settings
display. The
Mode (T3100)
display will appear.
Normal mode
Touch the
Normal mode
field to switch normal operating mode on. When normal
mode is in use the
Normal mode
field is shown highlighted.
Practice mode
Practice mode enables the user to practice and demonstrate program functions
available on the X-ray unit. In practice mode no radiation is generated when the
exposure button is pressed and there is no connection to a computer. The C-arm will
move normally but no radiation will be generated and no radiation warning signals will
be given, i.e. this is a “dummy run” function for training and demonstration purposes.
For example, users might want to demonstrate the C-arm movements before taking
exposures of children or nervous patients.
Touch the
Practice mode
field to switch practice mode on. When practice mode is in
use the
Practice mode
field is shown highlighted and symbols indicating that radiation
and PC connection are disabled will appear in the status bar on all displays.
To switch practice mode off, touch the
Normal mode
field.

Table of Contents

Other manuals for Planmeca ProOne

Related product manuals