Chapter A: GENERAL & TECHNICAL DATA 9 EMC INFORMATION
Technical manual Planmeca ProOne 23
9 EMC INFORMATION
Use of any accessories and cables other than those specified in Planmeca ProOne
X-ray unit’s documentation, with the exception of cables sold by Planmeca as
replacement parts for internal components, may result in increased emission or
decreased immunity of the X-ray unit.
Planmeca ProOne X-ray unit should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the Planmeca ProOne X-ray
unit should be observed to verify normal operation in the configuration in which it is
used.
Guidance and manufacturer’s declaration - electromagnetic emissions
Planmeca ProOne X-ray unit is intended for use in the electromagnetic environment specified
below. The customer or the user of the Planmeca ProOne X-ray unit should make sure that it is
used in such environment.
Emissions Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
Planmeca ProOne X-ray unit uses RF energy
only for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
Planmeca ProOne X-ray unit is suitable for use
in all establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies