3 PROGRAM SETTINGS Chapter B: SETTINGS
46 Planmeca ProOne Technical manual
3.2 Program: Program presets (P2200)
To adjust program default settings, first select the option Program presets (P2200) on the
Program settings
display. The
Program presets (P2200)
display will appear.
Estimated DAP
Touch the
View estimated DAP
field to set the X-ray unit so that the estimated Dose
Area Product (DAP) value will be shown on the main display. The DAP value
indicates the highest radiation dose the patient will be exposed to during the
exposure.
The
View estimated DAP
field is highlighted when the function is switched on. To
switch off the function, touch the
View estimated DAP
field again.
Default settings for programs
DEC target value: Touch the arrow buttons in the
DEC target value
field to adjust the
setting for DEC (Dynamic Exposure Control) target value. DEC target value is
expressed in percentage in comparison to DEC calibration value. The setting can be
adjusted between 50% (lower exposure values -> brighter image) and 200% (higher
exposure values -> darker image). The recommended setting is 100% (default value).
NOTE
DEC target value will be shown on the
Program presets (P2200)
display only if DEC has
been activated on the
Program licences (P2100)
display.
Lateral TMJ default angle: Touch the arrow buttons in the
Lateral TMJ default angle
field to adjust the setting. The factory default imaging angle for lateral exposures is 17
degrees.
TMJ 2/2 shift: Touch the arrow buttons in the
TMJ 2/2 shift
field to adjust the setting
for the automatic shift. The setting determines the imaging position for the second
(2/2) exposure in Double lateral and Double PA TMJ programs. The shift can be
adjusted between -15 mm and 0 mm. The factory default setting is -10 mm, i.e. the
imaging position will move 10 mm forward for the open jaw exposure (2/2).