9 EMC INFORMATION Chapter A: GENERAL & TECHNICAL DATA
26 Planmeca ProOne Technical manual
Recommended separation distances between
portable and mobile RF communications equipment and Planmeca ProOne X-ray unit
Planmeca ProOne X-ray unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of Planmeca ProOne X-ray unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the Planmeca ProOne X-ray unit as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output power
of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2.5 GHz
0.01 0.2 0.2 0.3
0.1 0.4 0.4 0.7
1 1.2 1.2 2.4
10 4.0 4.0 8.0
100 12.0 12.0 24.0
d 1,2 P=