E02-012 Lift motor Limit sensors The lift motor limit sensors (upper and
lower) are either defective or disconnected.
Check the lift motor limit sensors.
E02-013 Short circuit There is a short circuit in the lift motor.
Check the lift motor cable.
E02-015 Collimator
movement
The collimator is not functioning correctly.
Check that the mechanism is not stuck. The
collimator may need to be serviced.
E02-016 Temple support
motor
Limit sensors The temple support motor limit sensor
(open or close) is either defective or
disconnected.
Check the temple support motor limit
sensor.