GX-6000 Operator’s ManualReplacing the H
2
S Scrubber Disk and the Charcoal Filter • 118
4 . To replace the H
2
S scrubber disk, remove the old one from its recess in
the sensor gasket and replace it with the new one making sure that it is
properly seated. You do not have to remove the sensor gasket to replace
the H
2
S scrubber disk. It sits in a recess on top of it.
5 . To replace the charcoal filter, pull the old one out of its recess in the
sensor gasket if it did not already come out and replace it with a new one.
You do not have to remove the sensor gasket to do this. The charcoal
filter sits in a recess on top of it.
Figure 23: Replacing the H
2
S Scrubber Disk and Charcoal Filter