EasyManuals Logo

RKI Instruments GX-6000 User Manual

RKI Instruments GX-6000
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #182 background imageLoading...
Page #182 background image
173 • Using Maintenance Mode GX-6000 Operator’s Manual
WARNING: When the GX-6000 is restored to its default configuration,
the zero and span values for each channel are reset. You must
recalibrate all active channels if you restore the GX-6000 to
its default configuration.
1 . From the main menu, place the cursor in front of RESTORE
DEFAULT.
2 . Press and release POWER/ENTER. The Restore All? Screen appears
asking if you want to restore the default configuration.
3 . If you do not want to restore the default configuration, press and release
DISP/LOCK to return to the main menu.
If you do want to restore the default configuration, continue with Step 4.
4 . Press and release the POWER/ENTER button. A screen appears asking
you to confirm that you want to restore the default configuration.
5 . Press and release the POWER/ENTER button. The screen will indicate
that the default configuration has been restored and return to the main
menu.
MAINTE
RESTORE DEFAULT
RESTORE ALL ?
YES : ENTER
NO : DISP
MAINTE
RESTORE DEFAULT
ARE YOU SURE ?
YES : ENTER
NO : DISP

Table of Contents

Other manuals for RKI Instruments GX-6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the RKI Instruments GX-6000 and is the answer not in the manual?

RKI Instruments GX-6000 Specifications

General IconGeneral
BrandRKI Instruments
ModelGX-6000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals