243 Calibrating the 10.0 eV/Benzene PID Sensor GX-6000 Operator’s Manual
5 . Press and release the POWER/ENTER button to enter the TUBE CAL
Menu.
6 . Use the SHIFT (PANIC) button to place the cursor next to AUTO
CAL.
7 . Press and release the POWER/ENTER button to enter the TUBE CAL
Auto Calibration Menu.
8 . To view or change the benzene concentration to be used for the TUBE
CAL:
a. Use the SHIFT (PANIC) button to place the cursor next to
CONCENTRATION.
TUBE CAL
GAS CALIBRATION
AUTO CAL
SINGLE CAL
ESCAPE
TUBE CAL
GAS CALIBRATION
CONCENTRATION
START CAL
ESCAPE