245 Calibrating the 10.0 eV/Benzene PID Sensor GX-6000 Operator’s Manual
15 . If the calibration fails, a failure screen will appear and the instrument
will go into alarm.
Press and release the RESET button to clear the alarm and return to the
TUBE CAL Auto Calibration Menu. See “Troubleshooting” on
page 104.
16 . Use the SHIFT (PANIC) button to place the cursor next to ESCAPE,
then press and release the POWER/ENTER button to return to the TUBE
CAL Menu.
17 . Use the SHIFT (PANIC) button to place the cursor next to ESCAPE,
then press and release the POWER/ENTER button. You will be
prompted to remove the filter tube and tube holder.
18 . Remove the tube holder from the GX-6000’s inlet fitting.
19 . Press and release the POWER/ENTER button. You will return to the
Calibration Mode Menu.
20 . Use the SHIFT (PANIC) button to place the cursor next to the
NORMAL MODE menu option, then press and release the POWER/
ENTER button to return to Measuring Mode.
21 . Detach the calibration tubing from the tube holder’s inlet fitting.
22 . Unscrew the demand flow regulator from the calibration cylinder.
TUBE CAL
BNZ
FAIL
ppm
TUBE CAL
REMOVE HOLDER
YES :ENTER
NO :DISP