5. Make the settings relevant to your network:
Enter the IP address of the monitor (last three digits must be
different from the PC address)
Enter netmask (must match the netmask of the PC)
Enter default gateway (if available)
The symbol is displayed on the monitor.
The LAN connection is set up.
NOTE
Settings you make in this dialog window are active directly. You do
not need to save or confirm them.
6. Continue based on your device variant:
For devices with bioimpedance measurement, continue at step 7.
For devices without bioimpedance measurement, continue at step
10.
7. Ensure that the measuring mat is correctly suspended in the magnetic
catch of the monitor.
8. In the Peripherals tab, press the WiFi button.
9. In the WiFi mode drop-down menu, press the desired mode:
WiFi direct (BIA mat): The monitor and the measuring mat
communicate with one another via a direct WiFi connection
WiFi off: The monitor and the measuring mat communicate with
one another via the infrared interface