78 • secamVSA 535/seca mBCA 525
17-10-05-353-002g_10-2020 B
6. Use the keypad of the measuring mat for the following steps:
a) Enter patient position: Press the relevant position button
b) Press the button: Start the measurement procedure
7. Observe measuring progress on the measuring mat:
8. Remove the electrode cables from the electrodes.
9. Remove the electrodes from the patient.
10. Suspend the measuring mat in the magnetic catch of the monitor
➔ Suspending the measuring mat in the magnetic catch.
The updated seca patient file is transmitted to the monitor.
The measurement procedure is complete when the Measurement
successful message appears.
You have the following options for continuing:
► Press the button: Add comment
► Press the button: Save measurement
► Press the Analysis tab: ➔ Analyzing measurement
► Press the Vital signs tab: ➔ Measuring vital signs
► Press the button: Discard measurement
Measuring mat
Measurement progress
Electrodes
Position
button
1. • Patient position is given
• Electrodes being tested
On On
2. • Measurement in progress
• Measuring results will be stored
on the measuring mat
Flashing On
3. Measurement end: Measuring
results are stored on the
measuring mat, waiting for
transmission to the monitor
Off Flashing