Send mode
D
eactivated: Data output is deactivated
On request: Data is output when there is an explicit request from a host computer
via TCP/IP using CoLa2
Continuous and on request: Data is output continuously via UDP to a defined target
address and also when there is an explicit request from a host computer via
TCP/IP using CoLa2
Data content
Measurement data: Distance data with reflector detection and RSSI
Field interruption: Data on the light beams in interrupted fields of the active monitor‐
ing case
Application data: Status of inputs and outputs that are used in the monitoring case
table
Device status: Information on the status of the safety laser scanner (e.g., cut-off
paths, errors)
Configuration of data output: Information on the angular range actually being used
(for technical reasons, data from a slightly larger angular range than the one set
may be output in some cases)
Angular range
You can define the range within which measurement data and data relating to field
interruptions is output.
Complementary information
For additional information on data output, see the technical information “microScan3:
outdoorScan3: Data output via UDP and TCP/IP” (part number 8022706).
7.15 Transfer
Transferring configuration
DANGER
H
azard due to lack of effectiveness of the protective device
Persons and parts of the body to be protected may not be recognized in case of non-
observance.
When transferring the configuration, the protective device’s existing configuration may
be overwritten.
b
Check the configuration carefully before transfer.
b
Make sure that the desired device is connected during transfer.
At first, the configuration only exists as a project, namely as a configuration file. The
conf
iguration must be transmitted to the device.
At the left, you see the values configured in the project for the device. If the device is
connected, you see the values saved in the device at the right.
The compatibility of the configuration is checked during transfer.
CONFIGURATION 7
8021913/15ZW/2019-11-14 | SICK O P E R A T I N G I N S T R U C T I O N S | microScan3 – EFI-pro
111
Subject to change without notice