Restarting safety function and connections
•
T
he device’s function is also re-established after serious faults if the cause has
been rectified.
•
Communication with the device is interrupted (connections for configuration,
safety function and data not relating to safety). The device sets up communication
again automatically after restarting.
•
Communication beyond the device is not impaired.
Restarting device completely
•
The device behaves exactly as it does when the voltage supply is switched off and
back on again.
•
The device’s function is also re-established after serious faults if the cause has
been rectified.
•
Communication with the device is interrupted (connections for configuration,
safety function and data not relating to safety).
•
Communication beyond the device is interrupted. This may also affect devices
which communicate beyond the device.
7.18.2 EtherNet/IP
If the safety laser scanner has already been connected to a control and should be con‐
nec
ted to another control, the link to the old control must be explicitly removed.
1. Click on Remove link to control (reset ownership) to remove the link to the control.
7.18.3 Factory settings
Before reconfiguring the device, you can reset all settings to factory settings.
Reseting safety function to factory settings
•
T
he configuration for the safety function is reset to factory settings.
•
Communication beyond the device is not impaired.
Reseting whole device to factory settings
•
The configuration for the safety function is reset to factory settings.
•
The configuration of device communication is reset to factory settings (connec‐
tions for configuration, safety function and data not relating to safety).
7.18.4 Managing passwords
Assigning or changing passwords
1.
Establish a connection to the device.
2. In the device window, under Service, choose the entry User password.
3. Choose the user group in the User password dialog box.
4. Enter the new password twice and use Accept to confirm.
5. If you are requested to log in, log in as an Authorized customer.
✓
The new password is valid for the user group immediately.
Resetting a password
If you have forgotten a password, you can reset it.
1. Request the form for resetting your password from SICK support.
2. Connect to the device in Safety Designer.
3. In the device window, under Service, choose the entry User password.
4. Choose the Reset password option in the User password dialog box.
CONFIGURATION 7
8021913/15ZW/2019-11-14 | SICK O P E R A T I N G I N S T R U C T I O N S | microScan3 – EFI-pro
115