Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
250 Operating Instructions, 06/2018, A5E36281317-AC
Totalizers (2.14.)
8.6.14.1
Transducer selector (2.14.1.)
Sets the transducer index for all parameters applicable to this sub-menu.
Running Totalizers (2.14.)
Use these features to view, reset, or preset the eight-digit display totalizer when Sensor
Mode (2.1.3.) (Page 163
) is set for
Flow rate in open channel
or
. The eight-
digit totalizer is divided into two groups of four digits. The four most significant totalizer digits
are stored in
Running Totalizer High (2.14.2.) (Page 250), and the four least significant digits
are stored in Running Totalizer Low (2.14.3.) (Page 251). Adjust these values separately to
set a new total.
Running Totalizer High (2.14.2.) (Page 250) = 0017
Running Totalizer Low (2.14.3.) (Page 251) = 6.294
= 00176.294
Totalizer units are dependent upon programming. Enter zero 0 (if required) to reset the
totalizer to zero. Alternatively, enter any other (applicable) value, to preset the totalizer to the
necessary value.
In RUN mode, the full totalizer value can be viewed in Measurement View 2 in Auxiliary
Reading (see The LCD Display (Page 51)).
Note
A second point is ava
ilable only if the dual-point feature is enabled.
Running totalizers high (2.14.2.)
Shows and/or alters the four most significant digits of the totalizer value.
• Running totalizers low (2.14.3.) (Page 251)
• Totalizer decimal position (2.14.4.) (Page 251)
• Totalizer multiplier (2.14.5.) (Page 252)