Parameter reference
8.6 Setup (2.)
HydroRanger 200 HMI
Operating Instructions, 06/2018, A5E36281317-AC
165
8.6.1.6
Material (2.1.6.)
Specifies material type.
Index
Single-point model
Dual-point model
Global
Transducer
Options
*Liquid
Solid
Alters
TVT type (2.11.3.6.) (Page 230)
8.6.1.7
Short shot frequency (2.1.7.)
Adjusts the short shot transmit pulse frequency, in kHz.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 41 to 46 kHz, nearest acceptable value is returned
Altered by
Transducer (2.1.5.) (Page 164)
Related
Number of short shots (2.1.11.) (Page 166)
Long shot duration (2.1.10.) (Page 166)
Short shot bias (2.11.2.10.) (Page 225)
Short shot floor (2.11.2.11.) (Page 226)
Short shot range (2.11.2.12.) (Page 226)
8.6.1.8
Long shot frequency (2.1.8.)
Adjusts the long shot transmit pulse frequency, in kHz.
Index
Single-point model
Dual-point model
Global
Transducer
Values
Range: 41 to 46 kHz, nearest acceptable value is returned
Altered by
Transducer (2.1.5.) (Page 164)
Related
Number of long shots (2.1.12.) (Page 167)
Short shot frequency (2.1.7.) (Page 165)
Short shot duration (2.1.9.) (Page 166)