© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Replacing boards and components 109
Page 109 of 148
For internal use only
x Configure the following parameters in User Setup for MEMOSKOP:
Language Set for language of the country.
Display of realtime clock Configure Yes. Select No only upon customer request.
Display of rotated images Configure Yes.
Monitor zoom Configure No.
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number Configure 100.
Disable dose rate selection
in user setup
Set to "No" if equipped with an dose-area product measure-
ment device. The menu "Storage of Dose on disk" and the
menu "Dose on Hardcopy output" are visible in User Setup.
Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL. Do not
enter the leading zero.
Product name Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes. Depending on country-specific regulations
configure Yes or No.
Create external disk Configure No.
Default LUT Configure 2. Program a different LUT curve only upon cus-
tomer request.
Default Edge Configure 1. Program a different Edge enhancement value
only upon customer request.
Software backward com-
patible
Configure No.
Operating mode parame-
ters
Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: configure to both monitors.
Split screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date / Time Set the current date and time.
Hospital Name Enter the hospital name.