112 Replacing boards and components
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 112 of 148
For internal use only
Number of Monitors Configure 1 for SIREMOBIL with 1 monitor. Configure 2 for
SIREMOBIL with 2 monitors.
Hardcopy Check whether 5000 is programmed in "4 Max Hardcopy
time". If not, program this value.
Edge 1 coef set Check whether 3 is programmed; if not, program this value.
Edge 2 coef set Check whether 8 is programmed, If not, program this value.
Motion detection setup Check whether the following are programmed "MD1:
Threshold" 100 and for "MD2: Threshold" 120. If not, pro-
gram these values.
Date / Time format Set the country-specific format.
Keyboard Set to the keyboard layout being used.
Language Set for language of the country.
Display of realtime clock Configure Yes. Select No only upon customer request.
Display of rotated images Configure Yes.
Monitor zoom Configure No.
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number Configure 700.
Disable dose rate selection
in user setup
Set to "No" if equipped with an dose-area product measure-
ment device. The menu "Storage of Dose on disk" and the
menu "Dose on Hardcopy output" are visible in User Setup.
Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL. Do not enter
the leading zero.
Default LUT Configure 2. Only program a different LUT curve upon cus-
tomer request.
Default Edge Configure 1. Program a different edge enhancement value
upon customer request only.
Software backward com-
patible
Configure No.
Product name Program "SIREMOBIL Iso-C".
Default program change-
able
Default: Yes. Depending on country-specific regulations
configure Yes or No.
Create external disk If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.