© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Replacing boards and components 113
Page 113 of 148
For internal use only
x Configure the following parameters in User Setup for MEMOSKOP:
x Perform the IQ quick test.
NOTE
With present MODULARIS system, the parameters for the
Litho program have to be set again according to the LITHOS-
TAR MODULARIS Service Instructions.
Enable Bridge If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.
Delay 3D-PC/ DICOM
Bridge
Only available if option "3D Reconstruction" is installed.
Configure 120.
Interface to 3D-PC Only available if option "3D Reconstruction" is installed.
Configure Yes.
Operating mode parame-
ters
Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: Configure to monitor A. For
SIREMOBIL with 2 monitors: Configure to both monitors.
Split screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date / Time Set the current date and time.
Hospital Name Enter the hospital name.
Storage of Dose on disk If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.
Dose on Hardcopy output If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.