© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Replacing boards and components 115
Page 115 of 148
For internal use only
Monitor zoom Configure No.
Monitor split Configure Yes.
Monitor text Configure Yes.
Maximum image number Configure 900.
Disable dose rate selection
in user setup
Set to "No" if equipped with an dose-area product mea-
surement device. The menu "Storage of Dose on disk" and
the menu "Dose on Hardcopy output" are visible in User
Setup. Otherwise set to "Yes".
Unit serial number Program the serial number of the SIREMOBIL. Do not
enter the leading zero.
Product name Program "SIREMOBIL Iso-C".
Software backward compati-
ble
Configure No.
Default program changeable Default: Yes. Depending on country-specific regulations
configure Yes or No.
Create external disk If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.
Enable Bridge If the option "3D Reconstruction" or the option "DICOM
Bridge" is installed or a MOD drive is present, configure
Yes. Otherwise configure No.
Delay 3D-PC/ DICOM Bridge Only available if option "3D Reconstruction" is installed.
Configure 120.
Interface to 3D-PC Only available if option "3D Reconstruction" is installed.
Configure Yes.
Default LUT Configure 2. Only program a different LUT curve upon cus-
tomer request.
Default Edge Configure 1. Program a different edge enhancement curve
upon customer request only.
Default LUT for Subtraction Configure 1. Only program a different LUT curve upon cus-
tomer request.
Subtraction K-Factor, phase
A
Program "32". Program a different value only upon cus-
tomer request.
Subtraction K-Factor, phase
B
Program "MD 2". Program a different value only upon cus-
tomer request.
Subtraction Time of phase A Program "4.5". Program a different value only upon cus-
tomer request.
Roadmap K-Factor, phase A Program "32". Program a different value only upon cus-
tomer request.