116 Replacing boards and components
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 116 of 148
For internal use only
x Configure the following parameters in User Setup for MEMOSKOP:
x Perform the IQ quick test.
NOTE
With present MODULARIS system, the parameters for the
Litho program have to be set again according to the LITHOS-
TAR MODULARIS Service Instructions.
Roadmap K-Factor, phase B Program "MD 2". Program a different value only upon cus-
tomer request.
Roadmap K-Factor, phase C Program "MD 2". Program a different value only upon cus-
tomer request.
Roadmap Time of phase A Program "4.5". Program a different value only upon cus-
tomer request.
Operating mode Parameters Change the programmed parameters only upon customer
request.
VCR output Configure to monitor A.
Hardcopy Release For SIREMOBIL with 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors: Configure to monitor B.
Text monitor For SIREMOBIL with 1 monitor: Configure to monitor A.
For SIREMOBIL with 2 monitors: configure to both moni-
tors.
Split screen Configure horizontally. Configure vertically only upon cus-
tomer request.
Display of time Configure Yes.
Set Date / Time Set the current date and time.
Hospital Name Enter the hospital name.
Storage of Dose on disk If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.
Dose on Hardcopy output If dose-area product measurement device is present (Dia-
mentor), configure Yes. Otherwise configure No.