© Siemens, 2008 SPR2-230.840.01.01.02 SIREMOBIL Iso-C
08.08 CS PS SP
Adjustments / Programs 75
Page 75 of 148
For internal use only
Display of rotated images Yes: When pressing the image rotation key, the current
image will be computerized and displayed. NO: The rotated
image will not be computerized when pressing the image
rotation key. For SIREMOBIL Iso-C always set to YES.
Product name Allows you to change the MEMOSKOP name displayed dur-
ing system booting. Enter SIREMOBIL Iso-C.
Diagnostics Allows you to call test routines for repair by the manufacturer
of the MEMOSKOP. The MEMOSKOP can be reset via
"Load factory default", as required. Subsequently, you have
to program accordingly. The same applies if you replace the
MEMOSKOP
Create External Disk (1) MEMOSKOP CE and MEMOSKOP CE-100: Program the
value to "No". All other types of MEMOSKOP: If a MOD drive
is implemented and/or the option "3D Reconstruction" is
installed and/ or the option DICOM Bridge is installed, pro-
gram the value to "Yes".
Enable Bridge (1) MEMOSKOP CE and MEMOSKOP CE-100: Program the
value to "No". All other types of MEMOSKOP: If a MOD drive
is implemented and/or the option "3D Reconstruction" is
installed and/ or the option DICOM Bridge is installed, pro-
gram the value to "Yes".
Delay 3D-PC/ DICOM Bridge
(1)
Only available if option "3D Reconstruction" is installed.
Configure 120.
Interface to 3D-PC Only available if option "3D Reconstruction" is installed.
Configure Yes.
Monitor zoom Yes: When pressing the key for MEMOSKOP - Zoom, the
image will be displayed enlarged. NO: The MEMOSKOP -
Zoom key has no function. For MEMOSKOP C-E set to NO.
Otherwise set to YES.
Monitor split Yes: Horizontal or vertical image split possible. No: No hori-
zontal or vertical image split possible. Program Yes.
Monitor text Yes: Superimposes text. No: No text superimposition. For
SIREMOBIL Iso-C always set to Yes.
Maximum no. of images The number of stored images can be programmed. Program
the value to 700 for MEMOSKOP C and to 900 for
MEMOSKOP SUB. For MEMOSKOP C-E 100 program the
value to 100. For MEMOSKOP C-E, this parameter is not
used.
Default programs changeable Default: Yes. The organ programs can be protected against
changes by setting "No".
Default LUT After switch-on, the programmed brightness/contrast curve
(LUT) is selected automatically: Default: LUT 2
Default LUT for Subtraction After switch-on, the programmed brightness/contrast survey
(LUT) is selected automatically for the subtraction mode,
when equipped with a MEMOSKOP C-Sub. Default: LUT 1
Default Edge After switch-on, the programmed edge enhancement is
selected automatically: For MEMOSKOP SUB, set to 1.
Software compatibility Not implemented. Must be set to "No".
Function: Comments: