76 Adjustments / Programs
SIREMOBIL Iso-C SPR2-230.840.01.01.02 © Siemens, 2008
08.08 CS PS SP
Page 76 of 148
For internal use only
Unit serial number Enter the serial number of the SIREMOBIL Iso-C when it is
equipped with a DICOM bridge. The serial number is
required to identify the images transferred. Please note that
you have to omit the leading 0 when entering the serial num-
ber.
Disable dose rate selection in
user setup
This means "Yes" for SIREMOBIL Iso-C without a dose-area
product measurement device. This means "No" for a SIRE-
MOBIL Iso-C with an dose-area product measurement
device.
Subtraction K-factor, phase A Noise suppression, phase A, is programmed in the SUB-
mode. Phase A: Mask acquisition.
Subtraction K-factor, phase B Noise suppression, phase B, is programmed in the
SUB-mode. Phase B: Acquisition of fill image.
Subtraction Time of phase A The time of phase A in the SUB-mode is programmed.
Phase A: Mask acquisition.
Roadmap K-factor, phase A Noise suppression, phase A, is programmed in the roadmap
mode. Phase A: Mask acquisition.
Roadmap K-factor, phase B Noise suppression, phase B, is programmed in the roadmap
mode. Phase B: Acquisition of fill images.
Roadmap K-factor, phase C Noise suppression, phase C, is programmed in the roadmap
mode. Phase C: Acquisition of roadmap images.
Roadmap Time of phase A The time of phase A in the roadmap mode is programmed.
Phase A: Mask acquisition.
(1) in menu Diagnostics
Function: Comments: