11.3 Electrical data
Devices for general use
Measured current
I
Measuring
(Pt 100) 0.3 ... 1.0 mA
I
Measuring
(Pt 1000) 0.1 ... 0.3 mA
Devices in explosion-protected version
Equipment protection by means of intrinsic safety
For connecting to circuits with the following
peak values
U
i
≤ 30 V
I
i
≤ 100 mA
P
i
= P
o
(transmitter)
C
i
= 700 pF/m
L
i
= 15 µH/m
Equipment protection by means of non incendive
For connecting to circuits with the following
peak values
U
n
= 30 V
U
max
= 32 V
1)
1)
Maximum safety voltage
Equipment protection by means of Explosionproofed / Dust-Ignition proofed
For connecting to circuits with the following
peak values
U
max
= 45 V
(U
max
= 35 V for USA/Canada)
P = 25/37/50/100 mW
Eective internal capacitance and internal inductance
Display SITRANS TH100 ∑
C
i
16 nF 13 nF 29 nF
L
i
3 µH 106 µH 109 µH
Technical data
11.3 Electrical data
SITRANS TS100/TS200/TS300/TS500/TSinsert/TSthermowell