8.3.24.1 Enabling button lock
Procedure
1. Navigate to the parameter view.
Navigating in the views (Page 60)
2. In the parameter view, select the "Button lock" parameter.
3. Press the
button.
The "EDIT" symbol 󹪞ashes.
4. Select ON with the or button.
5. Use the button to con󹪝rm.
Result
The display automatically returns to the measurement view.
The display automatically changes between the measured values every 12 seconds.
The symbol for button lock "LL" and the measured value ID are displayed alternately.
8.3.24.2 Disabling button lock
Procedure
To disable the button lock, press and hold the button for 5 seconds.
Result
The symbol for Button lock "LL" is hidden.
You can operate the device using the buttons.
Parameter assignment
8.3 Parameter assignment over device with display
SITRANS TS100/TS200/TS300/TS500/TSinsert/TSthermowell
94 Operating Instructions, 08/2020, A5E47810090-AA