4. Using a 1/8” allen wrench remove four socket flat countersunk head cap screws (P/N
0004-346-000) (L) and separate the actuator strike plate (P/N 0702-014-111) (B) by
lifting it upwards.
Caution: the extended hall sensor (P/N 0702-014-119) (H), the rover hall sensor (P/N
0702-014-106) (E), the hall sensor (P/N 0702-014-906) (J), and the docker power coupler
assembly (P/N 0702-014-114) (F) are all mounted to the strike plate. Each component has
a wiring harness attached to it. Be careful not to get the connectors caught on anything
when the removing the actuator strike plate (P/N 0702-014-111) (B).
5. Using a 1/8” allen wrench, remove two socket flat countersunk head cap screws
(P/N 0004-346-000) (L).
6. Remove the docker power coupler assembly (P/N 0702-014-114) (F) and set aside.
4-7