4. Push the button next to the OK icon to select the desired REPROGRAMMING
menu option. Push the button next to the ESC icon to return to the previous
menu (see figure 3).
Figure 3: Select the Desired Option
5a. RESET EEPROM
Choosing this option will reset all of the values that are stored on the EEPROM.
After selecting RESET EEPROM, the system will display a warning to indicate
that the options will be erased (see figure 4). The values that will be restored to
system defaults are:
1. Minimum and maximum water flow
2. Docking odometer
3. Display and brightness settings changed by the user
4. Fluid level sensor calibration*
* This will require the technician to recalibrate the canisters as detailed in procedure 6.4 Performing the Canister Calibration