Stryker Neptune 2 - Analyzer Settings and Patient Leakage Current Limit for Rover

Stryker Neptune 2
232 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6. Measure and record the patient leakage in all combinations listed in table 6.5.2.
Analyzer Settings
Leakage Current Limit
Polarity Neutral
Normal Closed < 10 uA
Reverse Closed < 10 uA
Normal Open < 50 uA
Reverse Open < 50 uA
Table 6.5.2 Analyzer Settings and Patient Leakage Current Limit for Rover
6-21