Procedure 6.4 – Performing the Canister Calibration
The canister calibration must be performed when volume sensing components
are replaced. These components include the small canister (P/N 0702-001-300), large
canister (P/N 0702-001-320), main controller circuit board (P/N 0702-001-800), level
sensor assembly (P/N 0702-001-880), and float sensors (P/N 0700-001-155). The
technician will need an accurate device capable of measuring 1 liter of fluid. A suitable
beaker is listed in Appendix A: List of Materials.
1. To access the System Settings Menu, hold down the display button in the
upper-right hand corner and dock the unit. (See figure 1)
Figure 1: Accessing Menu While Docked