Fluid Diverter Assembly P/N 0702-001-350
Caution: When removing elbow connectors be aware of any water and
detergent that may be present!
1. Remove the coupling clip (P/N 702-1-204) (C) from the coupling block (P/N 702-1-
600) that holds the diverter inlet water hose assembly (P/N 702-1-341) (B) and
remove the hose from the block.
2. Remove the coupling clip (P/N 702-1-204) that holds the large tank inlet water hose
assembly (0702-001-340) (C) to the large canister cap assembly (0702-001-220)
(AH) and remove the hose from the cap. (Cap will be labeled Water Port)