17. With a manifold inserted into the small canister manifold receptacle, suction
the 1 liter prepared in step 16. Make sure as much fluid as possible is inside
the canister and that as little as possible remains in the suction tubing and
manifold. In order for the calibration to be accurate it is important that the
majority of the measured fluid is inside the tank. If only calibrating the small
canister, proceed to step 20.
18. Fill a 1 liter beaker with warm water and record the fluid amount. In most
cases the volume will not be exactly 1 liter. Make sure to record the exact
value within 1ml. (For example, 1002 ml) This will be used later on in step 27.
19. With a manifold inserted into the large canister manifold receptacle, suction
the 1 liter prepared in step 18. Make sure as much fluid as possible is inside
the canister and that as little as possible remains in the suction tubing and
manifold. In order for the calibration to be accurate it is important that the
majority of the measured fluid is inside the tank.
20. Turn off the suction by pressing the vacuum system power icon and
allow the fluid to stabilize in the canister(s) for one minute.
21. Turn off the main power switch. Access the System Settings Menu by holding
down the display button in the upper-right hand corner while turning on the
main power switch. (See figure 6)
Figure 6: Access the System Settings Menu
6-15