6. Initially the values displayed on the screen will be rapidly changing. Once the
screen displays DIAG, the values displayed for each magnet will slowly stop
changing and eventually settle at one number.
7. Determine which value is not in the correct range as discussed in the overview.
8. Unplug the rover and replace the faulty component.
6-42