Sening
®
is a registered trademark of FMC Technologies
111
MN F09 002 US  DOK-383E  Issue/Rev. 3.65 (02/24)
MultiFlow  W&M Calibration
After the access rights (master) have been checked a security inquiry
follows:
Data transmitted from MultiFlow to chip card?”
Press <F2> (“No) and the procedure is stopped with no change to the
chip card.
Confirm with <F1> (“Yes”).
Should the chip card have never been used (or only used for other
purposes) there is a further warning:
Chip card error! Incorrect format. Continue?”
Press <F2> (“No) and the procedure is stopped with no change to the
chip card.
Confirm with <F1> (“Yes”).
In view of the memory quantity needed, a CS-CC-512 chip card is to be
used.
During the transfer, which lasts approx. 90 seconds, the following
message appears:
Transfer in progress... Please wait”
The following message is shown at the end of the transmission:
Successful data transmission!
Confirm this message with <F1> (“OK”).
The parameters have now been transmitted.
Any error is accompanied by an appropriate notification and the
transmission needs to be repeated.
Deposit the chip card in the provided holder inside the MultiFlow housing.