8
W&M Calibration
8.1 Calibrating the measuring system
The MultiFlow supports the calibration of the measuring system using a
special routine (service menu). Before calibration you must define the
basic parameters of the product (<dg_ref_source_inline>Verweis --->
Produktdefinition</dg_ref_source_inline>):
•
W&M code, parameter 3.5.n.n.3
•
Product name, parameter 3.5.n.n.1
•
Product type, parameter 3.5.n.n.2 (2 = liquid product)
•
Volume unit, parameter 3.5.n.n.5 (1 = liters)
•
Date and time, parameter 1.2 (for the log)
The marker n.n determines the product or register number of the product
that is to be calibrated.
After the calibration function has been selected (in the SERVICE menu
4.2 "Calibration”), the calibration accuracy is automatically selected (ten
times the resolution for standard delivery).
☺ The normal delivery mask appears (see below) in which the product to be
calibrated is selected.
Preset Volume, if required
Function Key Assignment
Normally, entry of the preset volume is not used.
In this case the discharge is terminated with the <Stop> key when the
calibration vessel is full.