9
Configuration of the MultiFlow
The main menu is reached via function key <F3> in the start screen, and
enables the configuration of the MultiFlow. It contains the product
parameters as well as settings for the display, printer and the sensors.
Selection menu by pressing the
<F3> key.
The menu item "8 Autom. residue removal" is available from version 5.00
and higher (approval according to MID).
All entries in the menu can be viewed almost without any restriction.
Changes, however, are only possible with restrictions, as in particular for
the parameter list various access rights have been defined, such as
'Driver', 'Master' or 'Calibrate'.
Only settings in the 'Driver' group can be changed at any time. Changes
for all other settings will trigger a security query.
Parameters in the groups 'Master' and 'Calibrate' require master access,
i.e. the operator must identify himself as a master by entering the
appropriate password.
With erroneous entry of the password (master code), an entry is made
in the log book. Each time the password is entered incorrectly, the
waiting period before a new entry becomes possible is extended (by 1,
15, 60 min).
When parameters in the 'Calibrate' group are changed, the non-
resettable event counter is automatically incremented. Please also refer
to chapter 8.2 "Securing the settings on a chip card (MID version
only)" / page 110, Electronic W & M seal (soft seal).