112
BEFORE REQUESTING SERVICE
"LAN restruct error" Failed to reset the LAN connection. Confirm that the LAN cable
connection and the LAN setting are in the correct way.
"Failed to get IP address." Failed in IP address auto assignment. Set a fixed IP address, or
check if the DHCP server is running.
"Unknown Error" Displayed in case of a LAN error other than the LAN errors men-
tioned previously.
"IP address conflict" Displayed when the IP address is duplicated.
"Previous measurements
are left. Please press the
Clear button."
Displayed when the output of all output-set data fails.
"Output not set" Displayed when all output settings are OFF.
"RS232C DATAOUT" RS232C data output is in process.
"RS232C SUCCESS" RS232C data output is completed.
"RS232C FAIL" RS232C data transmission failed.
"Check eyelid" Displayed when the patient's eyelid covers the pupil. Tell the
patient to open their eye as wide as possible.
"Please touch pupil on the
screen"
Displayed when the XY position is not aligned. Touch the pupil and
bring the screen to the pupil center.
"Please focus" Displayed when the measuring nozzle position is too far. Press the
forward button and adjust the measuring nozzle position.
"Nozzle at limit position" Warns that the Z motor position has reached the nozzle limit.
"TOO CLOSE" Warns that the measuring head is too close to the patient's eye.
"TOO FAR" Warns that the measuring head is too far from the patient's eye.
"Please check the
DATE/TIME"
The battery for the built-in clock has become worn out. Before
using, confirm the time and date on the SETUP menu.
If the message comes up frequently, call your service engineer.
"Chinrest Error" Displayed when the chinrest is not correctly connected or not con-
nected at all.
"Turn off the unit?" Displayed to confirm whether to move the chinrest and measuring
head to their last positions.
"Perform air check?" Displayed to confirm whether to perform air check.
"Air checked. Press OK to
continue."
Displayed when normal air check operation is confirmed.
"Confirm abnormal
action(-) of air check"
Displayed when an anomaly occurred during air check. Check the
measuring nozzle for any foreign matter.
"Confirm abnormal
action(+) of air check"
Displayed when an anomaly occurred during air check. Check the
measuring nozzle for any foreign matter.
"Range of Input value is
2 - 10"
Displayed when the measurement count is set out of the setting
range. Set it within the input range.
"Please wait until packing
mode is finished..."
Indicates that the packing operation is in process. Wait until it is
completed.