BEFORE REQUESTING SERVICE
"Packing mode is fin-
ished.Please turn off the
device."
Indicates that the packing operation is completed. Switch off the
machine.
"First Octet is 1 - 223
Range"
Displayed when the first octet of IP address, default gateway, pri-
mary DNS server or secondary DNS server is set to a value out of
the specified input range.
"Value is irregular.
Input valid value"
Displayed when the octet is set to a value out of the specified input
range. Enter a value within the input range.
"The IP address is 0 - 255
Range"
Displayed when the subnet mask is set to a value off the input rule.
Enter a value within the input rule.
"At least 3 characters are
required for operator ID
prefix."
Displayed when the input examiner ID prefix is less than 3 charac-
ters. Enter a prefix with 3 characters.
"Incorrect password" Displayed when the password inputted to select a special mode is
incorrect.
"Please don't place the
face on the chinrest, to pre-
pare for next measure-
ment."
Displayed to confirm whether to place the face on the chinrest or
not, because the measuring head shall move to the "stand by
mode" position of the initial setting.