4-18 Acquiring Data
14. On the MassLynx Tune window DRE tab, specify the values of Mass and
Magnification factor in the table.
Tip: The ideal magnification factor is 40 and you should aim for values
between 20 and 60. The magnification factor must always be greater
than 10.
Mass values and example magnification factor results for a mixture of
chloropentafluorobenzene and heptacosa in CI negative ion mode:
Mass Magnification factor
201.9609 50.65
451.9743 43.28
513.9711 35.25
594.9663 28.04
632.9632 28.11