Performance specifications F-3
For example:
In EI positive ionization mode, less than 10 ng of methyl stearate (m/z 298.3)
injected onto a capillary column will give, for all peaks above mass 70 and
greater than 10% of the base peak intensity, 1.25 mDa RMS error or better,
with the 4.0 GHz TDC.
Dynamic range
Enabled by dynamic range enhancement (DRE) and digital dead time
correction (DDTC), the dynamic range of the instrument is over four orders of
magnitude.