EasyManua.ls Logo

Waters GCT Premier - Inserting the DCI Probe into the Insertion Lock

Default Icon
382 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8-4 The Direct Chemical Ionization (DCI) Probe
To load sample into the DCI probe:
1. Push the DCI tip onto the end of the DCI probe.
2. Dissolve the sample in a volatile GC-grade solvent such as methanol to a
concentration of 1 ng/µL.
Tip: If necessary, to increase the amount of sample introduced via the
tip, use a correspondingly stronger solution. It is common to run samples
of between 1 ng and 1 µg.
3. Deposit 1 µL (1 ng of sample) on the end of the DCI probe tip using a
syringe.
Tip: The droplet stays at the tip due to surface tension.
4. Wait for the solvent to evaporate.
5. Insert the probe into the source.
See also: Inserting the DCI probe into the insertion lock on page 8-4.
Inserting the DCI probe into the insertion lock
To insert the loaded probe into the ion source:
1. Ensure that the instrument is pumped-down (evacuated) and is ready
for use.
See also: Chapter A, Starting-up and Shutting-down the Instrument.
2. Ensure that the probe insertion lock valve is closed.
Warning: To avoid contamination from samples, wear
chemical-resistant, powder-free gloves at all times when
handling the probe.
Caution: The wire filament at the end of the DCI probe tip is
fragile. Be careful not to damage the tip when handling it.
Caution: Damage to the instrument might occur if the probe insertion
lock is operated with the probe incorrectly positioned.

Table of Contents

Related product manuals