2-6 Preparing the instrument for EI/CI operation
Changing the EI/CI outer source assembly
Removing the EI/CI outer source from the instrument
To remove the EI/CI outer source:
1. On the MassLynx Tune Window, click Vacuum > Vent Source.
2. Wait for the Vacuum Status LED to display steady yellow.
3. If fitted, remove the GC column from the instrument.
See also: Installing the GC interface on page 6-3.
4. Move the GC away from the right side of the instrument.
5. Remove the inner source from the outer source.
See also: Removing the EI/CI inner source on page 2-3 and Removing
the FI/FD inner source on page 9-13.
Warning: The source component scan be contaminated with
toxic and biohazardous materials. Wear chemical-resistant,
powder-free gloves at all times while handling the
components.
Warning: The source is hot and can cause burn injuries to
the skin. Allow the source to cool before handling it.
Warning: To avoid burn injuries, allow the GC column and
interface to cool before handling.