B-42 The MassLynx Tune window
Analyser Solvent Trip Changing this value changes the position of the
green trip level indicator in the analyser vacuum
display.
The function shuts off all analyser and source
voltages if tripped and protects the instrument
against pressure surges. Set a value of about 5.5,
which corresponds to about 3e-6 mbar.
Source Turbo The speed at which the source turbo operates.
When the source is pumped down (evacuated) the
value should be between 95 and 100%. A value
below 90% indicates a cooling fault.
See also: Contacting Waters on page 11-14.
Analyser Turbo The speed at which the analyser turbo operates.
When the analyzer is pumped down (evacuated)
the value should be between 95 and 100%. A
value below 90% indicates a cooling fault.
See also: Contacting Waters on page 11-14.
Engineer tab buttons
Button Description
Open/Close Isolation
Valve
Controls the position of the isolation valve
between the source and the analyser housing,
regardless of whether the instrument is in
standby or operate mode.
Diagnostic Message Log Displays an updated list of diagnostic messages.
Engineer tab parameters: (Continued)
Parameter Description