17-3
IM 701240-01E
17
Troubleshooting, Maintenance, and Inspection
Errors in Execution (700 to 799)
Code Message and Corrective Action Section
701 Fine copy is not possible when waveform acquisition is in progress. Press the START/STOP 7.1, 12.1
key to stop the waveform acquisition first.
703 Undo is not possible since data that existed immediately before initialization and auto setup 4.4, 4.5
is not available.
704 Can not be executed while running. Press START/STOP key to stop acquisition. 7.1
707 Can not Start while data out. Wait until output is completed. 7.1
708 Can not data out while running. Press START/STOP key to stop acquisition. 7.1
710 File not found. Check the file. Chapter 13
711 Cannot manipulate files while image printing is in progress. Wait until image printing is complete. Chapter 13
712 Can not compress this hardcopy image. Turn off compression switch. 13.12
713 Calibration failure. Disconnect the input and execute again. If it fails again, servicing is necessary. 4.6
714 Temperature module calibration failed. Calibration value is invalid. Check the input signal. 4.6
716 Realtime recording to the internal hard disk is valid when the sampling rate is slower than the 7.5
values shown below.
1~2 CH : 100kS/s, 3CH : 50kS/s,
4~5CH : 20kS/s, 6~11CH : 10kS/s,
12~18CH : 5kS/s
717 Too many channels to realtime record at the current record time. 7.5
Decrease the number of channels by turning them OFF.
718 Too many channels to realtime record at the current record length. –
Turn OFF channels to reduce the number of channels or shorten the record length.
721 Can not operate while data out. Wait until output is completed. 7.5
723 Cannot load the realtime record. 7.5,13.15
724 Balancing failed. –
Failed CH: *
* indicates the channel number of the channel for which automatic balancing failed.
725 Cannot be executed. Acquisition not active. 7.1
727 Insufficient output data. Increase Mag or widen the Time Range interval. 12.1
728 Image is being printed. Abort or wait until printing is complete. –
732 Cannot be executed while computation is in progress. Chapter 10
733 Failed to measure statistics. Waveform data may be missing. If Cycle Statistics is specified, the 11.6
instrument may be configured in a way that fails to detect the cycle.
735 Executing file Load, Save, or Format. Abort or wait until it is complete. –
736 Image is being printed or saved. Abort the operation or wait until the execution of the –
command is complete.
737 A setup of an action trigger can’t be carried out under the state of on. 6.18
739 Cannot be executed when realtime record or print setting is ON. 7.5
740 A time base can’t be carried out under the state of the outside clock. 5.20
742 Because there are too many channels, it can’t start in the length of the present record. 5.2,
Appendix 1
743 The indication mode of the history can’t be carried out except for “Average”. 11.1
744 Average practice can’t be done because the record length of the history exceeds the record 11.1
length that it can be carried out.
745 Set the trigger mode to Auto or Log for realtime hard disk recording. If the number of actions 6.1, 7.5
is set to Continuous, the only valid trigger mode is Log.
748 Cannot start at the current record length. 5.2, 6.1
Shorten the record length or meet the following condition.
• Set the trigger mode to Auto and increase the record time to longer than 1 s to enable roll mode.
• Set the trigger mode to Single or Log.
749 Averaging mode is not possible when the trigger mode is Single, SingleN, or Log. 6.1
Change the trigger mode.
17.2 Messages and Corrective Actions