17-4
IM 701240-01E
Code Message and Corrective Action Section
753 Cannot start when the HISTORY menu is being displayed. 11.2
Press the ESC key to clear the menu before starting.
755 Time base cannot be set to external clock when set to envelope or box average. 5.20, 7.2
756 Cannot set accumulate during roll mode display. Turn Off accumulate. 5.2, 8.4
757 Cannot be executed when the acquisition mode is set to average. Change the mode. 7.2
758 Cannot copy if 50 pages is exceeded. Change the multiplier or range. 12.1
759 Realtime recording to the internal hard disk is valid when the record length is longer than 501 k. 5.2, 7.5
760 Cannot set the current drive on a realtime partition using communication commands. 7.5
761 Cannot be executed on realtime recorded waveforms. 7.5
764 File recorded in realtime is currently being analyzed. Files being analyzed cannot be deleted. 7.5, 13.16
765 File recorded in realtime is currently being analyzed. 7.5, 13.18
The name of files being analyzed cannot be changed.
766 File recorded in realtime is currently being analyzed. 7.5, 13.6
Partitions containing files that are being analyzed cannot be formatted.
767 128 and more file can’t be formed in the real-time area. 7.5
768 The file which failed in the real-time record can’t be read.
769 Cannot start Single(N) when accumulate is On. Change the accumulate to Off. 6.1, 8.4
773 The measuring range is up to 10 MWords from measure start (TimeRange1). 11.5
774 Synchronizing signal not detected. 6.17
775 Set acquisition mode to Normal when using a wave window trigger. 6.17
776 The wave window trigger cannot be used if the sampling rate is faster than 500 kS/s or slower 6.17
than 10 kS/s. If the record length is shorter than 25 k, set the record time longer than 100 ms.
777 Range over. Change to an appropriate range then retry shunt calibration. 5.15
778 Statistical processing cannot be performed on waveforms recorded in real time. 11.6
779 Cannot detect a effective input signal at a Sync. channel. 6.17
780 Firmware was not overwritten in the following slots, since the version of the firmware in the module
and that of the replacement firmware were the same. Check the versions on the overview screen.
SLOT: XX
781 Cannot start while the following setup menus are displayed. Press the ESC key to clear
the menu, and then start.
(ALL CH, ZOOM, MEASURE, CURSOR, and PrintSetup)
782 Firmware was not overwritten, since the version of the firmware in the printer and that 17.4
of the replacement firmware were the same.
783 Checksum error in the printer control program. 3.4
Recycle the power. If the error occurs again, maintenance service is required.
784 Failed to transmit the printer control program. 3.4
Recycle the power. If the error occurs again, maintenance service is required.
785 Failed to write the printer control program. 3.4
Recycle the power. If the error occurs again, maintenance service is required.
787 Cannot be carried out during recording. 7.1
Press the START/STOP key to stop the waveform acquisition first.
788 Cannot start while the alert dialog box is open.
Press the ESC key to clear the dialog box, and then start.
789 Cannot copy if 100 pages is exceeded. 9.7, 12.1
Change the multiplier or range.
790 Fine Print or Zoom Print is not possible when X-Y display is present. 8.6, 12.1
791 Up to 30 items can be registered. 4.2
792 Not entered in “Yomi” or “Goku.” 4.2
793 Cannot specify characters other than “Hiragana” for “Yomi.” 4.2
794 Kanji registration error. Below are the possible causes. 4.2
More than 17 items are registered with the same “Yomi.”
Exceeded the dictionary registration capacity.
17.2 Messages and Corrective Actions