17-6
IM 701240-01E
Code Message and Corrective Action Section
873 It is an unacceptable parameter to set up to the present module. Chapter 5
875 Cannot be set to a range of 200 sec to 30 day during roll display. –
876 Because a record length is too long, it can’t be set up by the present number of indication channels. 5.2,
Appendix 1
877 Cannot be set because there are too many display channels at the current record length. 5.2
Shorten the record length.
878 Zooming is not available when the number of displayed points of the FFT waveform is less 10.3
than 50 in the Zoom window.
880 Cannot change this setting during realtime recording. 7.5
881 The time base cannot be set to external clock when the acquisition mode is set to envelope 7.2, 5.20
or box average.
882 Spaces are not allowed for trace labels. 8.10
883 Cannot set accumulate during roll mode display. 5.2, 8.4
884 Cannot change the History parameter when accumulate is ON. Turn OFF accumulate first. 8.4
885 P-P compression cannot be used to save when a record length is 1 K. 13.8
887 A sampling speed was too fast, and it could be folded, and a real-time queue was needed. 7.5
Slow down a sampling speed.
888 A sampling speed was too fast, and it could be folded, and a real-time buffer was needed. 7.5
Slow down a sampling speed.
889 Cannot set On this module. 6.17
890 Settings can not be entered for channels on which no strain module is mounted. –
891 Cannot be changed when Variable is turned ON. –
895 Cannot be set when the trace is set to a frequency module. –
System Operation Errors (900 to 999)
Code Message and Corrective Action Section
901 Failed to backup setup data. Will initialize. Backup battery may be low. 7.7
902 System RAM failure. Maintenance service is required. –
903 System ROM failure. Maintenance service is required. –
906 Fan stopped. Power off immediately. Maintenance service is required. –
907 Backup battery is flat. Maintenance service is required to replace the back-up battery. 3.4
908 Internal temperature is too high. Power off immediately. Maintenance service is required. 3.4
909 Illegal SUM value. Maintenance service is required. –
910 RAM read/write error. Maintenance service is required. –
911 Memory bus error. Maintenance service is required. –
912 Communication driver error. –
914 Time out occurs in Communication. –
915 EEPROM read error. EEPROM may be damaged. Maintenance service is required. –
916 EEPROM write error. EEPROM may be damaged. Maintenance service is required. –
917 No module installed. Install the module. 3.3
918 Turn ON the internal hard disk motor. 13.3
919 The current module installation condition and the setup data are inconsistent. –
Will initialize.
922 Cannot back up the acquisition memory. Memory will be initialized. 7.7
The battery for acquisition memory backup may be flat.
928 Battery for waveform memory is flat. 7.7
Waveform memory was initialized.
929 MS bus error occurred. –
930 Slot 1 EEPROM error. –
931 Slot 2 EEPROM error. –
17.2 Messages and Corrective Actions