AutoTest Sytem Operation
Subject to Export Control, see Cover Page for details.
5-17
Config Commands (cont)
set_audio_sinad_llimit [ limit dB ]
Sets lower limit for afsinad meter.
Format: Decimal limit value in dB
set_audio_sinad_llen [ state 0/1 ]
Enables/Disables lower limit check on afsinad meter.
Valid states are:
0 = Disabled
1 = Enabled
set_audio_hn_llimit [ limit dB ]
Sets lower limit on fm deviation meter with rms deviation type.
Format: Decimal limit value in dB
set_audio_hn_llen [ state 0/1 ]
Enables/Disables the lower limit check on hum and noise meter.
Valid states are:
0 = Disabled
1 = Enabled
set_audio_hn_average
Sets number of averages used to calculate measurement.
Format: Whole number
Range: 1 to 250
set_audio_hn_lockref [ state 0/1 ]
Enables/Disables the reference lock on hum and noise meter.
Valid states are:
0 = Disabled
1 = Enabled
set_audio_snr_llimit [ limit dB ]
Sets lower limit on fm deviation meter with rms deviation type.
Format: Decimal limit value in dB
set_audio_snr_llen [ state 0/1 ]
Enables/Disables the lower limit check on hum and noise meter.
Valid states are:
0 = Disabled
1 = Enabled
set_audio_snr_average
Sets number of averages used to calculate measurement.
Format: Whole number
Range: 1 to 250
set_audio_snr_mode [ state 0/1 ]
Sets type of SNR measurement to be performed.
Valid states are:
0 = Hum and Noise
1 = Normal Signal to Noise Ratio