EasyManuals Logo

Agilent Technologies 6850 Series User Manual

Agilent Technologies 6850 Series
337 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #280 background imageLoading...
Page #280 background image
Troubleshooting
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 281
9. If the 1/8-inch copper split vent tubing was
disconnected from the inlet, confirm that it
has been re-attached without leaks.
10. The EPC module can fail in a mode where the
inlet EPC valve does not open, causing this
symptom.
An electronic leak detector is very useful in
isolating large leaks. Order Agilent product
number 5182-9646 (120 volts) or 5182-9648
(220 volts).
Case 2: Actual inlet pressure increases
slowly and does not reach the setpoint
NOTE: Check all of the parameters from Case 1
before proceeding.
1. Set the inlet to be tested to the following
configuration:
Confirm that the installed column is con-
figured correctly to the capillary inlet
and that there is only one column config-
ured to the inlet
Split mode
Inlet pressure 25 psi
Split flow 400 mL/min
Confirm that the actual inlet pressure is
still lagging the setpoint

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6850 Series and is the answer not in the manual?

Agilent Technologies 6850 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model6850 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals