Appendix C Message Lists
C-4
Table C-1 Error Message List (Cont'd)
Message Phenomenon
Operation prohibited while Ext.Trig. displayed
This operation cannot be performed while the
external synchronized measurement is in
Operation prohibited while Power Monitor
displayed
This operation cannot be performed while the
power monitor measurement is in progress.
An abnormality occurred in the optical
attenuator at start-up.
Optical input power too high
Insert attenuator or decrease input level.
The optical input level is too high. Use an optical
attenuator or reduce the optical level.
Optical power too low for Optical Unit
auto-adjustment. Adjust input level.
The optical input level is too low for accurate
optical axis adjustment.
Optical power too low to calibrate wavelength
Adjust input level.
The optical input level is too low for wavelength
calibration.
Optical Unit failed alignment adjustment
Optical axis adjustment cannot be performed
during start-up.
Optical Unit failed auto alignment
An abnormality occurred at optical axis auto
adjustment at start-up.
Optical Unit failed calibration data test
Contact Anritsu or representative.
There was a fault in the calibration data at
start-up.
Optical Unit failed FPGA data test
Contact Anritsu or representative.
There was a fault in the FPGA data at start-up.
Optical Unit failed grating control
There was a fault in the grating control at
start-up.
Optical Unit failed memory test at boot
There was a fault in the memory test at start-up.
Optical Unit failed offset adjustment
Check no optical power input.
Offset adjustment failed at start-up. Check that
there is no optical input.
Optical Unit failed program test
Contact Anritsu or representative.
An abnormality was detected in the program test
at start-up.
There is an error in option xxx.
Optional light source error
An abnormality was detected in the light source
option at start-up.
Pase enabled only when PLZN Nulling set for
Method.
Pase was selected at Write to when a method
other than PLZN Nulling was selected at the
Peak->Level disabled when Calculated set for
Trace Type
Change Trace Type to setting other than
Peak → Level processing is not supported when
Trace Type is set to Calculate.
Change Trace Type to something other than
The recall processing was completed.
Reading of the file is completed.
Resolution bandwidth calibration failed
The resolution calibration is failed.
Resolution Cal in progress
The resolution calibration is in progress.
The save processing was completed.
Saving of the file is completed.